  universal ac input low leakage current 0.18ma protections: short circuit / overload / over voltage / over temperature fixed switching frequency at 100khz free air convection for 140w and forced air convection for 200w ul60601-1 medical safety approved with power good and fail signal output built-in remote on-off control built-in remote sens function model number output volts output amps(25cfm) ripple & noise convection line reg load reg volt tolerance single output mps200-3.3 3.3 volts(dc) 40 amps 80mv pk-pk 28 amps 0.5% 1.0% 2.0% mps200-5 5 volts(dc) 40 amps 80mv pk-pk 28 amps 0.5% 1.0% 2.0% mps200-12 12 volts(dc) 16.7 amps 100mv pk-pk 11.7 amps 0.5% 1.0% 2.0% mps200-15 15 volts(dc) 13.4 amps 100mv pk-pk 9.4 amps 0.5% 1.0% 2.0% mps200-24 24 volts(dc) 8.4 amps 150mv pk-pk 5.9 amps 0.5% 1.0% 1.0% MPS200-48 48 volts(dc) 4.2 amps 2000mv pk-pk 3.0 amps 0.5% 1.0% 1.0% dual output mpd-200a 5 volts(dc) 20 amps 80mv pk-pk 15 amps 0.5% 1.0% 2.0% 12 volts(dc) 8 amps 120mv pk-pk 5.4 amps 2.0% 4.0% +8,-5% mpd-200b 5 volts(dc) 20 amps 80mv pk-pk 15 amps 0.5% 1.0% 2.0% 24 volts(dc) 4 amps 180mv pk-pk 2.7 amps 2.0% +4,-6% 6.5% 200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458
model number output volts output amps(25cfm) ripple & noise convection line reg load reg volt tolerance triple output mpt-200a 5 volts(dc) 20 amps 80mv pk-pk 15 amps 0.5% 1.0% 2.0% 12 volts(dc) 7.5 amps 120mv pk-pk 5.0 amps 1.0% 5.0% 8.0% -5 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% mpt-200b 5 volts(dc) 20 amps 80mv pk-pk 15 amps 0.5% 1.0% 2.0% 12 volts(dc) 6.0 amps 120mv pk-pk 4.4 amps 1.0% 5.0% 8.0% -12volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% mpt-200c 5 volts(dc) 20 amps 80mv pk-pk 15 amps 0.5% 1.0% 2.0% 15 volts(dc) 4.7 amps 150mv pk-pk 3.3 amps 1.0% 5.0% 8.0% -15 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% mpt-200d 5 volts(dc) 20 amps 80mv pk-pk 15 amps 0.5% 1.0% 2.0% 24 volts(dc) 3.0 amps 180mv pk-pk 2.2 amps 1.0% 5.0% 8.0% 12 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% quad output mpq-200b 5 volts(dc) 15 amps 80mv pk-pk 12 amps 0.5% 1.0% 2.0% 12 volts(dc) 7.0 amps 120mv pk-pk 5.3 amps 1.0% 5.0% 8.0% -5 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% -12 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% mpq-200c 5 volts(dc) 15 amps 80mv pk-pk 12 amps 0.5% 1.0% 2.0% 15 volts(dc) 5.0 amps 150mv pk-pk 4.0 amps 1.0% 5.0% 6.0% -5 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% -15 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% mpq-200d 5 volts(dc) 15 amps 80mv pk-pk 12 amps 0.5% 1.0% 2.0% 24volts(dc) 3.0 amps 180mv pk-pk 2.3 amps 1.0% 5.0% 8.0% 12 volts(dc) 2.0 amps 300mv pk-pk 1.0 amps 0.5% 1.0% 5.0% -12 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% mpq-200e 5 volts(dc) 15 amps 80mv pk-pk 12 amps 0.5% 1.0% 2.0% 24 volts(dc) 2.7 amps 180mv pk-pk 2.1 amps 1.0% 5.0% 8.0% 15 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% -15 volts(dc) 2.0 amps 80mv pk-pk 1.0 amps 0.5% 1.0% 5.0% 200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458
input specifications input voltage range 90-264vac / 127-370 volts(dc) frequency range 47-440 hz inrush current, typ: cold start <25 amps @ 115vac input 40 amps @ 230vac input current 3.5 amps max @ 115vac 1.6 amps max @ 230vac power factor @ fl pf>0.95/230vac pf>0.98/115vac leakage current heart leakage < 0.18ma / 264vac patient leakage < 0.1ma / 264vac output specifications voltage and current see selection chart line regulation see selection chart load regulation see selection chart voltage tolerance (note 2) see selection chart ripple/noise (note 1) see selection chart hold up time @ fl 16ms @ 230vac 16ms @ 115vac setup, rise time @ fl: 1000ms, 20ms / 230vac 3000ms, 20ms / 115vac over temperature protection tsw1 temp 95c, shutdown auto recover over voltage protection 5volts only:115~135%, shutdown over current protection hiccup type 120~160% auto recover peak load (note 3) 120% of output min. min load single o/p 0% multi o/p 20% dc volt adj -5%~10% rated output volt by potential meter general specifications safety ul60601-1 / ul60950-1 tuv en60601-1 / tuv en60950-1 i e c 6 0 6 0 1 - 1 a p p r o v e d insulation resistance 100m? emi compliance to en55011, en55022 class b harmonic current compliance to en61000-3-2,-3 effciency 81% typ. isolation 4000vac input - output 1500vac input - ground 1500vac output- ground ems compliance to en61000-4-2,3,4,5,6,8,11; env50204 remote control rc+/rc-: 0~0.8vots(dc) = power on 4~10volts(dc) = power off power good/fail 500ms/1ms environmental specifications oper. temperature -20c to +70c (see derate curve) storage temperature -40c to +85c, 10~90% rh relative humidity 20 to +90% rh temperature coeffcient 0.04% / c (0-50c) mtbf 262.1khrs min, mil-hdbk-217f(25c) vibration 10~500hz, 2g10min./1cycle, period for 60min. each along x, y, z axes physical specifications size millimeters 177.8 x 107.2 x 35.5 inches 7 x 4.22 x 1.4 weight 23.28 oz (660g) note 1. ripple and noise are measured at 20mhz of bandwidth by using a 12 twisted pair-wire terminated with a 0.1uf & 47 uf parallel capacitor . 2. tolerance: includes set up tolerance, line regulation and load regula - tion. 3. 33% duty cycle max within every 30s. average output power should not exceed the rated power. 4. derating may be needed under low inpt voltages. please check the derating curve for more details. all specifcations are typical at nominal input, full load, and 25c unless other - wise noted 200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458
200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458
200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458
200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458
200w 1~4 output for medical type mps/d/t/q200 series www.astrodynetdi.com astrodyne usa: 1-800-823-8082 astrodyne pacific: 886-2-26983458


▲Up To Search▲   

 
Price & Availability of MPS200-48

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X